Experimental design study of calibration meter with digital potential difference meter
FU Shi-min
Abstract:The operation trouble , time -consuming and system error of box -type potentiometer are ex-pounded.These problems can be solved if the digital potentiometer is used .UJ33D type digital potentiometer can occur and digital direct voltage measurements;Measurement range is 0-2500.0 mV, the accuracy level is 0.01, the minimum resolution is 0.01 mV.The principle of calibrating the electric meter with digital potentiometer ,The values of resistance R1 , R2 and Rs were studied, and the experimental design was analyzed .
Keywords:digital potential difference meterbox potential difference meterexperimental design studycalibration voltage metercalibration current meter
Publication Date:2016-01-01
Online Publishing Date:2025-08-15(First online date of this platform, not the publication date of the document)
Pages:4( 223-226 )

ISTIC
ISSN:1673-0569
Year, Vol.(Issue):2016,37(3)