Preliminary Evaluation of the Performance of the Chemiluminescent Immunoassay System by CLSI File
ZHANG Bang-lin
GUO Yong-can
Abstract:Objective:To evaluate the performance of Beckman Access2 automated chemiluminescent immunoassay system in order to know its status. Method:Using the CLSI document EP10-2A for preliminary evaluation methods to approve the high,medium and low fresh serum of patients which were adopted as samples and was assayed FT3 in Beckman Access2 testing system.Result:The factors of carryover and drift,system deviation,imprecision,cross contamination in the ballpark,they had no effect on precision.Conclusion:The performance of Beckman Access2 automated chemiluminescent system is reliable and suitable for clinical laboratory to use.
Keywords:Chemiluminescent systemImmunoassayFree triiodothyronine