Unaided and Aided ANL in Patients with Steep Drop Type of Hearing Loss
LIN Ying
CHEN Hanchao
ZHANG Guojun
Abstract:Objective The purpose of this study was to compare ANL values under unaided and aided conditions in subjects with steep drop type of hearing loss, as a measure of the effect of hearing aids. Methods Twenty subjects who had used hearing aids for over 6 months for steep drop type of hearing loss were selected. ANL values were measured under unaided and aided conditions, and the differences compared and analyzed using the SPSS19.0 software. Results①Unaided MCL (80.70 ± 8.05 dB HL) and BNL (73.75 ± 7.16 dB HL;) were higher than aided MCL and BNL (65.80 ± 9.03 dB HL and 59.25±6.93 dB HL, respectively) (P<0.001), while unaided and aided ANL values (6.95±4.20 dB HL vs 6.55 ± 4.21 dB HL) were not significantly different (P>0.05);②Real-ear analysis, which evaluates hearing aids ef-fectiveness, and ANL value showed a weak negative correlation (R=-0.2). Conclusions The ANL value may be used to predict effectiveness of hearing aids and guide hearing aid fitting in patients with steep drop type of hearing loss.
Keywords:ANLReal-ear TestEffect Evaluation
Publication Date:2017-01-01
Online Publishing Date:2025-08-15(First online date of this platform, not the publication date of the document)
Pages:4( 523-526 )
Chinese Journal of Otology

Chinese Journal of Otology

PKUISTIC
ISSN:1672-2922
Year, Vol.(Issue):2017,15(5)