Portable extracorporeal life support system in vitro and in vivo experiment
Liu Yu
Zhou Yuan-yuan
Yin Zong-tao
Li Hong-yi
Wang Hui-shan
Abstract:Objective To explore the performance of a portable extracorporeal life support (PELS) system in vitro and in vivo.Methods In vitro,PELS system was tested for stable assessment;in vivo,12 female swine were randomized into experiment group (n=6) and control group (n=6).Cardiopulmonary bypass (CPB) was repectively established with PELS and RotaFlow with 4 L/min of flow for 6 h.The plasma free hemoglobin and platelet activation were evaluated before initial of CPB,every 2 hs in CPB and after terminated CPB.Results In vitro,experiments confirmed that the stability of PELS system was better than or equal to Maquet company Rotaflow system;in vivo,the plasma free hemoglobin was lower than 20 g/L and the platelet activation was lower than 10% after 6 h performance of PELS system.Conclusion The portable extracorporeal life support system conducts the reliable performance in vitro and in vivo.
Keywords:PortableExtracorporeal life supportIn vivoIn vitro
Publication Date:2017-01-01
Online Publishing Date:2025-08-15(First online date of this platform, not the publication date of the document)
Pages:4( 38-41 )
Chinese Journal of Extracorporeal Circulation

Chinese Journal of Extracorporeal Circulation

ISTIC
ISSN:1672-1403
Year, Vol.(Issue):2017,15(1)