The value of SAFIRE in coronary computed tomography angiography with low iodine intake and low radiation dose
Abstract:Objective To investigate the image quality and radiation dose of CT coronary angiography (CCTA) using snogram affirmed iterative reconstruction ( SAFIRE) with lower kVp and 25%iodine intake reduction compared with standard dose CTA u-sing filtered back-projection (FBP).Methods 58 patients were randomly divided into two groups (29 for each):Group A, 100 kVp, 210 mgI/kg, and reconstructed with SAFIRE;group B,120 kVp, 280 mgI/kg and reconstructed with FBP .The CT attenua-tion values of aortic root (AR), left main coronary artery (LMA), right main coronary artery (RMA) , left circumflex artery (LCX) and background noise were measured.The CT dose index-volume (CTDI), Signal-to-noise ratio (SNR), and contrast-to-noise ratio ( CNR) were measured and statistically analysed .Image quality was evaluated by two senior radiologists using a quali-tative five-point scale.Results Compared with group B, group A significantly reduced the iodine intake (A:15.0 ±1.54g, B:19.7 ±3.0g, P <0.01) and radiation dose of group A was reduced by about 48.5% ( A:6.45 ±1.07mSv, B:12.53 ±2.27 mSv, P <0.01).CT attenuation value ,SNR and CNR of all segments had no significant differences ( P >0.05).Subjective image quality of two groups had no significant differences ( P >0.05).Conclusion Using SAFIRE under the condition of lower kvp and 25%iodine reduction could obtain satisfactory images meeting the diagnostic requirements .
Keywords:SAFIRELow radiationLow iodine intakeCoronary artery
Publication Date:2017-01-01
Online Publishing Date:2025-08-15(First online date of this platform, not the publication date of the document)
Pages:4( 52-55 )