Optimization of the noise index setting in chest computed tomography scan with automatic tube current modulation and adaptive statistical iterative reconstruction
LU Rui
HE Jian
ZHOU Kefeng
QIN Guochu
HU Anning
YANG Shangwen
ZHU Bin
Abstract:Objective To study the effect of noise index (NI) in the chest computed tomography (CT) image quality with au-tomatic tube current modulation scanning and adaptive statistical iterative reconstruction ( ASIR) , and to explore the most appro-priate NI to reduce the radiation dose .Methods A total of 193 consecutive patients were prospectively enrolled .They underwent chest CT scan with the application of automatic tube current modulation and ASIR (50%weight).Those patients were randomly divided into 6 groups with different NI settings (15, 18, 21, 24, 27 and 30) and were further divided into 3 subgroups according to their body mass index (BMI):underweight and normal (BMI≤23.9), overweight (24≤BMI≤27.9) and obese (BMI≥28) subgroups.Two radiologists made the subjective score of the image quality blindly and independently (well, <13;critical, 13~15;unacceptable, >15).Image noise (fat), signal to noise ratio (SNR, aorta vs.fat), contrast to noise ratio (CNR, aor-ta vs.fat) were measured on both lung and medium reconstruction images .CT Dose Index volume ( CTDIvol), Dose Length Product (DLP) and Effective Dose (ED) of each scan were recorded.Additionally, the percentage of images obtained with a tube current lower than 50 mA among the total image slices in each patient was recorded .Results With the increase of the NI , the subjective score increased ( image quality decreased ) , the SNR and CNR decreased .The subjective score of chest CT images were lower than or equal to 15 with all NI settings, while CT images of 4 patients (4/34) were scored 13~15 when NI=30.The sub-jective score and SNR of chest CT images with NI =30 was significantly higher and lower than those with NI =15~27 respectively ( P <0.05).The radiation dose of scanning with NI =24~30 was significantly lower than that with NI =15~21 ( P <0.05). When NI was set at 27 or 30, the percentage of images obtained with a tube current lower than 50 mA among the total image slices decreased with the increase of BMI ( P <0.05).When NI was set at 30, the radiation dose significantly increased with the in-crease of BMI ( P <0.05).Conclusion The NI of chest CT scan can be set at 30 with the application of automatic tube current modulation and 50%ASIR to reduce the radiation dose dramatically with reserved image quality .
Keywords:ChestLow doseAdaptive statistical iterative reconstructionAutomatic tube current modulationTomographyX-ray computed
Publication Date:2016-01-01
Online Publishing Date:2025-08-15(First online date of this platform, not the publication date of the document)
Pages:5( 1012-1016 )