Factors affecting the accuracy of thermal conductivity testing of graphite extruded board
WANG Xin
WANG Gongzhen
ZHOU Nannan
Abstract:This paper analyzes the accuracy of testing the thermal conductivity of plates using the steady-state guarded hot plate method from four aspects:equipment stability,human factors,laboratory environmental conditions,and the quality of the plates themselves.The results show that thermal conductivity instruments should be calibrated regularly to maintain test stability and im-prove the accuracy of test results;the thickness of the sample is directly proportional to the thermal conductivity—when the thick-ness increases or decreases by 1mm,the thermal conductivity change rate ranges from 2.86%to 5.31%,so the thickness should be measured multiple times to reduce measurement errors;changes in the temperature and humidity environment of the laboratory will affect the accuracy of thermal conductivity measurement of graphite extruded boards—for every 10℃increase in the laboratory temperature,the measured thermal conductivity value increases by 5.6%,and for every 10%increase in humidity,the measured thermal conductivity value increases by 3.1%;the thermal conductivity of graphite extruded boards is affected by aging time,and it changes with aging time,first increasing and then tending to be stable.
Keywords:insulation materialsthermal conductivitysteady-state guarded hot plate methodtesting accuracyinfluencing factors
Publication Date:2026-01-25
Online Publishing Date:2026-03-17(First online date of this platform, not the publication date of the document)
Pages:4( 87-90 )