Finite element analysis of fixed platform prosthesis and rotating platform prosthesis in total knee replacement
Zhang Li-chao
Zhang Li-min
Wu Li-zhu
Abstract:BACKGROUND:There are studies on three-dimensional finite element analysis of total knee replacement, but few research concerns contact pressure and contact area inside and outside polyethylene liner of fixed prosthesis and rotating platform prosthesis in total knee replacement. OBJECTIVE:To explore the contact pressure and contact area inside and outside polyethylene liner of fixed platform prosthesis and rotating platform prosthesis in total knee replacement. METHODS:We established three-dimensional finite element models of fixed platform prosthesis and rotating platform prosthesis in total knee replacement and compared contact pressure and contact area inside and outside polyethylene liner of fixed platform prosthesis and rotating platform prosthesis in total knee replacement. RESULTS AND CONCLUSION:The peak contact pressure inside and outside polyethylene liner of fixed platform prosthesis and rotating platform prosthesis in total knee replacement was minimum at 0° of genuflex. The peak contact pressure inside polyethylene liner was higher than that outside polyethylene liner of fixed platform prosthesis and rotating platform prosthesis in total knee replacement at 0°-90° of genuflex (P<0.05). The peak contact pressure inside and outside polyethylene liner of fixed platform prosthesis was higher than that of rotating platform prosthesis in total knee replacement at 60°-120° of genuflex (P<0.05). The contact area inside and outside polyethylene liner of fixed platform prosthesis and rotating platform prosthesis in total knee replacement was maximum at 0° of genuflex. The contact area inside polyethylene liner of fixed platform prosthesis in total knee replacement was lower than that of outside at 0°-30° of genuflex (P<0.05). The contact area inside polyethylene liner of fixed platform prosthesis in total knee replacement was lower than that of outside at 0°-90° of genuflex (P<0.05). The contact area inside and outside polyethylene liner of fixed platform prosthesis was lower than that of rotating platform prosthesis in total knee replacement at 30°-120° of genuflex (P<0.05). These results suggested that peak contact pressure inside polyethylene liner of fixed and rotating platform prostheses was higher than that outside in total knee replacement, and the inside contact area was lower than that of outside. Inside and outside contact pressure of the rotating platform prosthesis was lower than that of fixed platform prosthesis. Inside and outside contact area of the rotating platform prosthesis was higher than that of fixed platform prosthesis.
Machine-generated Keywords:
Publication Date:2016-01-01
Online Publishing Date:2025-08-15(First online date of this platform, not the publication date of the document)
Pages:6( 5801-5806 )