Production and functional testing of polyurethane valve of the pediatric Luo-Ye pump
Wu Yue-heng
Huang Huan-lei
Fan Rui-xin
Qi Zhou-cuo
Yao Li-ming
Wu Chang-li
Lu Ping-lan
Cheng An-heng
Xiao Xue-jun
Abstract: BACKGROUND:The mechanical valves used in the adult Luo-Ye pump have a large size and great destruction to blood, which are not suitable for infant ventricular assist pump. Therefore, designing and producing a high molecular valve with smal size and low incidence of thrombosis is a research hotspot. OBJECTIVE:To design and produce a valve of 20 mL infant Luo-Ye pump, and to test its basic functions and fatigue properties. METHODS:The size and shape of valve was designed with MASTERCAM software, polyurethane valve was obtained through producing the valve model and plastic injection;the static leakage, pressure drop and fatigue resistance of polyurethane valve were tested according to the ISO5840 requirements. RESULTS AND CONCLUSION:The polyurethane trefoil valve was produced, but the failure rate of plastic injection was high;the basic function of the trefoil valve met the ISO5840 requirements bascial y;after continuously operated 1.0×107 times, stroke volume of 20 mL Luo-Ye pump was changed 5.2%, and two polyurethane valves and valve leaflets did not change and damage. Polyurethane trefoil valve was designed and produced successful y;polyurethane valves could meet the needs of 20 mL Luo-Ye pump, which already have the ability to clinical trials.
Keywords:organ transplantationheart-lung transplantationventricular assist deviceLuo-Ye pumppolyurethanetrefoil valvefatigue resistancecardiac deviceNational Natural Science Foundation of China
Publication Date:2013-01-01
Online Publishing Date:2025-08-15(First online date of this platform, not the publication date of the document)
Pages:6( 5607-5612 )