Determination of ferulic acid in Bidouyan Oral Solution by HPLC
CAO Yun-fei
YE Fei
CHEN Na
Abstract:Objective To determine the content of ferulic acid by HPLC in Bidouyan Oral Solution. Methods The determination was carried out with Shimadzd C18 column. The mobile phase was methanol-1%acetic acid solution(30:70) and the flow rate was 1. 0 mL · min-1 . The column temperature was 30 ℃ and the detective wavelength was 321 nm. Results The content of ferulic acid can be determined by HPLC. The linear concentration range of these compound was 10. 22~102. 20 μg · mL-1 . The average recovery was 98. 84%. Conclusion The method was simple and accurate and can be used to control the quality of Bidouyan Oral Solution.
Keywords:Bidouyan Oral SolutionHPLCFerulic acid
Publication Date:2014-01-01
Online Publishing Date:2025-08-15(First online date of this platform, not the publication date of the document)
Pages:2( 691-692 )
Journal of Pharmaceutical Research

Journal of Pharmaceutical Research

ISSN:2095-5375
Year, Vol.(Issue):2014,(12)