Method Verification and Evaluation of Uncertainty for Determination of Heavy Metal Arsenic in Textiles by Atomic Fluorescence Spectrophotometry
Zhou Nan
Zhang Hongmei
Wang Liping
Chen Liang
Huang Long
Abstract:The content of arsenic in textiles was determined by atomic fluorescence spectroscopy.The linear range,detection limit,quantitation limit,accuracy and precision of the method were stud-ied and analyzed,and the uncertainty was evaluated.The results show that the detection limit of this method was 0.007 mg/kg,the quantitative limit was 0.017 mg/kg,the precision was less than 10%,and the relative standard uncertainty was 2.4%.The laboratory can meet the detection requirements of the standard method.
Keywords:atomic fluorescencetextilesarsenicmethod validationuncertainty
Publication Date:2024-04-20
Online Publishing Date:2026-05-22(First online date of this platform, not the publication date of the document)
Pages:4( 33-36 )