Development of Certified Reference Material for Total Phosphorus in Water
YU Ping
Abstract:In order to achieve the unity and accurate traceability of total phosphorus detection in water,and solve the problem of devia-tion between the results of different detection methods,this paper uses ammonium dihydrogen phosphate purity reference material as raw material,uses the weight capacity reference method to prepare 100mg/L total phosphorus solution reference material in water,and com-pletes the value verification by ion chromatography.The results show that the reference material has good uniformity and stability,and the accuracy of the measurement value is consistent with that of the national secondary reference material.The relative expanded uncer-tainty is 2.0%(k=2).It can be used for the calibration of total phosphorus detection instruments in water,the quality control of meas-urement process,and the confirmation and evaluation of analysis methods,providing reliable quantitative support for the accurate detec-tion of total phosphorus in water environment monitoring,drinking water supervision and other fields.
Keywords:total phosphorus in waterion chromatographcertified reference material(CRM)uncertainty
Publication Date:2026-09-30
Online Publishing Date:2026-11-18(First online date of this platform, not the publication date of the document)
Pages:5( 35-39 )