Research on Calibration Method of Battery Internal Resistance Tester Based on DC Discharge Method
CHEN Xiquan
LIN Ting
FANG Wei
Abstract:Aiming at the problem of insufficient calibration accuracy caused by dynamic changes in battery internal resistance,difficulty in calibrating standard values,and interference from battery polarization and contact resistance superposition in the traceability of DC discharge method battery internal resistance tester values,a calibration system integrating micro incremental measurement method is studied,and multi gradient incremental calibration experiments are conducted in combination with relevant standards.The results show that the relative error of this method within the range of(0~100)mΩ is less than 0.5%,the relative standard deviation of repeatability is less than 0.3%,and the comprehensive evaluation index of 20%~100%incremental gradient is less than 2%,which meets the technical requirements of the calibrated instrument's metrological characteristics and can effectively achieve high-precision value trace-ability,providing a standardized technical path for the calibration of such instruments.
Keywords:DC discharge methodbattery internal resistance testerincremental methodcalibration techniqueserror propagationmetro-logical traceability
Publication Date:2026-09-30
Online Publishing Date:2026-11-18(First online date of this platform, not the publication date of the document)
Pages:5( 30-34 )
Metrology & Measurement Technique

Metrology & Measurement Technique

ISSN:1004-6941
Year, Vol.(Issue):2026,52(9)