Study on the Influence of Temperature on Verification Results of Diaphragm Gas Meters and Environmental Control Measures
GUO Xiaowei
Abstract:As a mandatory verification measuring instrument for trade settlement,the measurement accuracy of membrane gas meters is crucial.This article,based on the calibration regulations of JJG 577-2012 "Membrane Gas Meter",deeply analyzes the influence of temperature on the calibration results from three dimensions:gas thermal properties,instrument mechanical structure,and standard de-vice stability.It proposes a comprehensive control strategy including the construction of a constant temperature and humidity environ-ment,air circulation homogenization,multi-point temperature control monitoring,and standardized operation,aiming to control the ambient temperature within the range of(20±2)℃.The experimental results show that the repeatability error of the verification after the implementation of this scheme is better than±0.3%,effectively improving the accuracy and credibility of membrane gas meter verification.
Keywords:diaphragm gas metermetrological verificationtemperature influenceindication errorenvironmental controlrepeatability
Publication Date:2026-08-30
Online Publishing Date:2026-11-18(First online date of this platform, not the publication date of the document)
Pages:4( 116-119 )
Metrology & Measurement Technique

Metrology & Measurement Technique

ISSN:1004-6941
Year, Vol.(Issue):2026,52(8)