Abstract:Transient methods for measuring the thermal conductivity of materials offer several advantages,including simple apparatus structure,short measurement time,a wide measurement range and small sample size requirements.Commercial instruments,primarily based on the transient plane source(TPS)method and the transient hot-wire(THW)method,have been widely used in materials re-search,production and inspection.However,owing to the relevant metrological technical specification system in China has not yet been fully established and the lack of well-established supporting calibration methods,the accuracy and reliability of measurement results obtained using transient thermal conductivity apparatus require further experimental evaluation.In this study,XPS,PMMA,quartz glass and 304 stainless steel were selected as reference samples.Effects of measurement power setting and ambient temperature,on the meas-urement performance of a DRE-C thermal conductivity apparatus were experimentally investigated and the resulting data were analyzed and discussed.
Keywords:transient thermal conductivity apparatusinfluencing factors of calibrationcertified reference materialexperimental analysis
Publication Date:2026-08-30
Online Publishing Date:2026-11-18(First online date of this platform, not the publication date of the document)
Pages:5( 111-115 )
Metrology & Measurement Technique