Analysis of the Influence of Different Mathematical Models on the Detection Results of Chemiluminescence Immunoassay Analyzers
HE Yi
QIU Guangyin
LI Yi
WANG Weiqi
SU Qiang
Abstract:The concentration calculation parameters of chemiluminescence immunoassay analyzer were obtained by establishing a math-ematical model between luminescence value and concentration.In this paper,exponential,four parameter,polynomial and linear models are used to fit the functional relationship between the two,and the fitting luminous value is calculated from the model parameters,and then the influence of each model on the detection results is evaluated from two aspects of goodness of fit and accuracy.The results show that polynomial,four parameter and linear models have better goodness of fit,while polynomial and four parameter models have better accuracy,especially in the low concentration region.
Keywords:mathematical modelgoodness of fitaccuracy
Publication Date:2026-07-30
Online Publishing Date:2026-11-18(First online date of this platform, not the publication date of the document)
Pages:3( 174-176 )
Metrology & Measurement Technique

Metrology & Measurement Technique

ISSN:1004-6941
Year, Vol.(Issue):2026,52(7)