Measurement and Quality Control of Output Power of Cardiac Radiofrequency Ablation Instrument
YI Dazhi
LIU Qing
WU Baoqiao
ZHANG Zhiqiang
CAO Yang
Abstract:In this paper,FLUKE QA-ES Ⅲ detector was used to test the repeatability of 10 power points of St.Jude Medical H700489 therapeutic instrument in the power control(PC)and temperature control(TC)modes,so as to analyze the quality control method of cardiac radiofrequency ablation instrument output power,and evaluate its uncertainty.The results show that in PC/TC mode,the relative error and repeatability of output power meet the technical requirements of industry standards,and the difference between the measured values of each power point is small;The uncertainty mainly comes from the measurement performance of the tester and the uncertainty of the superior measurement standard.This method can provide a scientific basis for the daily quality control of the output power of cardiac radiofrequency ablation instrument.
Keywords:radiofrequency ablationoutput powerpower measurementuncertaintyquality control
Publication Date:2026-07-30
Online Publishing Date:2026-11-18(First online date of this platform, not the publication date of the document)
Pages:4( 170-173 )