Establishment of the Reference Radiation Qualities for X-rays in the Medium-energy X-rays
WANG Jiawei
XU Jun
LIU Chunyan
ZHANG Zhicheng
Abstract:(80~250)kV medium-energy X-rays are widely used in radiological diagnosis and treatment.Establishing a reference radiation quality consistent with the international dose measurement reference system can ensure the comparability of measurement re-sults among standard laboratories and clinical users.This article determines the half-value layer by adjusting additional filtration and measuring attenuation ratio,establishes the reference radiation quality of medium-energy X-rays,and conducts experimental verifi-cation.The results show that the radiation quality of our laboratory is comparable to that of primary standard laboratories in terms of half-value layer,with a maximum deviation of-2.6%from the CCRI recommended value.
Keywords:medium-energy X-raysreference radiation qualityCCRIhalf-value layerradiological diagnosis and therapy