Development of the Second-class Standard Platinum Resistance Thermometer Handheld Measuring Instrument
YANG Yehua
HUANG Feng
CHEN Jia
SHI Liangxi
XIE Peng
YUAN Xiaodong
Abstract:The second-class standard platinum resistance thermometer needs to be equipped with electrical measuring instruments with high accuracy and high resolution.In this paper,a hand-held measuring instrument with high precision and low power consump-tion is designed based on four wire wiring and resistance comparison method.In order to obtain high-precision resistance value,the a-rithmetic mean filter is used to carry out high-order correction of ambient temperature and nonlinear high-order correction for the measured resistance value,and the specified reference function and deviation function are used to convert the measured resistance value into the corresponding temperature value.At the same time,the upper computer is developed based on labview2018 platform to realize wired and wireless reading and writing of key parameters of platinum resistance and measuring instrument,and downloading measure-ment data.After testing,the absolute error of the measuring instrument is less than±(0.002%r+0.2mΩ),and the uncertainty of re-sistance measurement is 0.009%.
Keywords:platinum resistance thermometermeasuring instrumenthigh precisionresistance
Publication Date:2026-07-30
Online Publishing Date:2026-11-18(First online date of this platform, not the publication date of the document)
Pages:5( 28-32 )