Research on the Design and Traceability Method of Flow Verification Device for Infusion Pump Detector
WU Huaiyi
Abstract:The reliability of infusion pump detectors is directly related to medical quality.To address the issue of traceability,this paper designs a high-precision verification device based on an active piston structure,tailored to meet the flow verification requirements.The device employs a weighing method and incorporates air buoyancy correction and evaporation suppression measures.Experimental verifi-cation was conducted within a flow range of(5~1000)mL/h.The results show that the maximum flow error of the device is 0.2468%and its stability is 0.1332%,both of which are superior to the±0.30%requirement specified in the JJG 1098-2014 regulation.The device exhibits good measurement accuracy and stability,providing a reliable means for quantity traceability.
Keywords:infusion pump testerverification deviceactive pistontraceability method