Evaluation of Uncertainty in Absorbed Dose in Water from Medical Electron Accelerator Radiation Source Using Monte Carlo Method
TANG Xun
ZHAO Shiyu
HAN Yuhua
SHI Shuguang
ZHANG Kunming
Abstract:To evaluate the measurement uncertainty of the absorbed dose in water for a 6mV X-ray beam from a medical electron ac-celerator,a TrueBeam accelerator was taken as the subject.Based on metrological procedures and international standards,the Monte Carlo method was employed to investigate the convergence of simulation times,the probability distribution of output quantities,and the impact of axial positioning errors ranging from(0.5~3.0)mm on uncertainty.The results indicated that the MCM simulation results were fully convergent when the simulation times exceeded 106;the probability distribution of output quantities approximated a normal distribution;the larger the positioning error,the higher the standard uncertainty of the absorbed dose in water,the wider the 95%proba-bility coverage interval,and the distribution tended towards a rectangular distribution.The relative uncertainty for a 3.0 mm positioning error was 1.25%,significantly affecting the determination of dose compliance.
Keywords:medical electron accelerator radiation sourceuncertainty evaluationmonte carlo methodabsorbed dose in water