Theoretical and Experimental Study on the Measurement Calibration Method and Temperature Distribution Characteristics of Tunnel-type Sterilization Dryers
ZHANG Guangzheng
SUI Pengxiao
HE Qirui
Abstract:As the core equipment in the production process of sterile preparations,the temperature distribution uniformity and steriliza-tion effect of the tunnel sterilization dryer are directly related to the quality and safety of drugs.However,there is currently a lack of cal-ibration standards for its specialized metrology in China.This article is based on the theory of dry heat sterilization,combined with the principles of thermodynamics and fluid mechanics,to construct a metrological calibration system and conduct experimental verifica-tion.The results showed that the equipment had good temperature uniformity in all areas under the operating condition of 320℃,with FH values higher than 1365 and endotoxin reduction of≥3 logarithmic orders,meeting the requirements of sterile formulation produc-tion.This can provide theoretical basis and practical guidance for its standardized calibration and optimized operation.
Keywords:tunnel sterilization dryerdry heat sterilizationtemperature distributionmeasurement calibrationthermal penetrationFH valueendotoxin challenge
Publication Date:2026-05-30
Online Publishing Date:2026-11-18(First online date of this platform, not the publication date of the document)
Pages:4( 187-189,194 )
Metrology & Measurement Technique

Metrology & Measurement Technique

ISSN:1004-6941
Year, Vol.(Issue):2026,52(5)