Quality Control Detection and Effect Analysis of Respiratory High-flow Humidifying Therapy Apparatuses
XU Lingyun
LI Fengjiao
CAO Aijie
Abstract:Based on the functional structure and clinical application characteristics of high flow respiratory humidification therapy instru-ment(HFNC),a quality control and detection scheme covering appearance,accessories,alarm function and core parameters(gas flow,oxygen concentration,gas temperature)was designed according to relevant specifications and standards,and a quality control system of"qualitative inspection+quantitative detection"and a quality control and detection worksheet of HFNC in hospital were construc-ted.Through the system property control detection of 30 different brands of HFNC in use in the hospital,this paper summarizes the types of hidden dangers of equipment functions,analyzes the causes of unqualified parameters and puts forward targeted improvement meas-ures,so as to provide technical support for the safe and effective clinical application of HFNC.
Keywords:respiratory high-flow humidifying therapy apparatusesquality controlparameter detectionequipment management
Publication Date:2026-05-30
Online Publishing Date:2026-11-18(First online date of this platform, not the publication date of the document)
Pages:5( 170-174 )