Research on Technical Renovation of Formaldehyde Gas Detector Calibration Devices
LI Xuehui
CHEN Xiyao
LI Yafei
Abstract:Aiming at the calibration device for formaldehyde gas detectors,this paper proposes a technical renovation solution based on the dynamic gas distribution tube method in accordance with the JJG 1022-2016 calibration regulations.By upgrading the dynamic gas distribution device and introducing cavity ring-down spectroscopy(CRDS)monitoring technology,the method achieves high-preci-sion continuous gas distribution and concentration monitoring of formaldehyde standard gases.The results show that the proposed solu-tion can reduce the uncertainty of the modified device from 4.0%(k=2)to 0.01 μmol/mol(low concentration range)and 2.8%(high concentration range),while optimizing the mixing cell temperature fluctuation from±0.2℃ to±0.02℃,providing reliable technical support for metrological traceability.
Keywords:formaldehyde detectordynamic gas mixingcavity ring-down spectroscopymetrological verificationtechnical transforma-tion
Publication Date:2026-05-30
Online Publishing Date:2026-11-18(First online date of this platform, not the publication date of the document)
Pages:5( 31-35 )
Metrology & Measurement Technique

Metrology & Measurement Technique

ISSN:1004-6941
Year, Vol.(Issue):2026,52(5)