Research on Angle Encoder Calibration Device and Method
TAN Dongxing
XIA Xue
TANG Xiaocong
ZHAO Yanlong
ZOU Congxia
Abstract:Aiming at the problems of insufficient accuracy and low efficiency in the existing calibration methods of angle encoder,this paper analyzes the influence of installation eccentricity and tilt error on measurement accuracy through theoretical modeling,and designs a calibration device integrating installation and fine-tuning mechanism.The modular chuck structure is adopted so that rotors of differ-ent sizes can be quickly clamped;The high-precision fine-tuning mechanism is adopted to ensure the automatic coaxial alignment be-tween the driving turntable and the shaft system of the angle encoder;The introduction of flexible clamping technology,using elastic ele-ments to buffer the relative runout between rotor and stator,reduces interference signals,and can realize the efficient and high-preci-sion calibration of absolute,incremental and other angle encoders.
Keywords:angle encoderinstallation eccentricityflexible clampingcalibration
Publication Date:2026-01-30
Online Publishing Date:2026-11-18(First online date of this platform, not the publication date of the document)
Pages:4( 61-64 )