Study on the Influence of Clamping Force of Water Meter Verification Device on Water Meter Error
WU Ershuai
WU Bin
Abstract:Based on the meter reading errors of water meters under different clamping forces in the verification device,this paper propo-ses corresponding improvement methods to address the impact of clamping force variations on the measurement results of turbine and ul-trasonic water meters,followed by experimental validation.The results indicate that the influence on ultrasonic water meters primarily stems from the deformation of rubber gaskets under higher clamping forces,which can be mitigated by replacing them with sealing gas-kets featuring reduced deformation.
Keywords:water meter verification deviceclamping forceinfluence
Publication Date:2026-01-30
Online Publishing Date:2026-11-18(First online date of this platform, not the publication date of the document)
Pages:3( 55-57 )
Metrology & Measurement Technique

Metrology & Measurement Technique

ISSN:1004-6941
Year, Vol.(Issue):2026,52(1)