Research on Calibration Method of Photogrammetric Scale Using Laser Interferometer
SUN Yanfeng
ZHANG Kun
Yu Qingbin
Niu Penglei
He Haijian
He Mingzhao
Abstract:To address the challenge of length traceability for photogrammetric scales in large-component measurement and overcome the precision limitations and systematic deviations of traditional calibration methods,this paper proposes a high-precision scale cali-bration approach.It uses laser interferometer displacement as the length standard and adopts the average displacement of multiple refer-ence points as the transfer carrier.Experimental results show that this method has an expanded uncertainty U=14pm(k=2),with normalized deviation En values less than 0.8 compared to the grayscale centroid method,demonstrating consistency.It offers high pre-cision,efficiency,and strong engineering applicability,providing an effective and practical calibration solution for photogrammetric scales.
Keywords:photogrammetryscale calibrationlaser interferometerlength traceabilitymulti-point averaging method
Publication Date:2026-01-30
Online Publishing Date:2026-11-18(First online date of this platform, not the publication date of the document)
Pages:4( 48-51 )