A biomechanics study of A-P Gamma nail
Abstract:Objective To study the complications and cuases of Gamma nail for the treatment of peritrochanteric fracture of femur by biomechanics test.Methods 16 fresh femoral specimens were divided into two groups, each consisting of eight specimens. One group was as control, another was as A-P Gamma nail. Evans typy Ⅰ intertrochanteric fracture were produced and fixed with A-P Gamma nial. In the standing position, press test were carried out on MST tested instrument for to compare the biomechanical property of A-P Gamma nail with the control.Results  When press was 300N, press strength of control was (1 568.06±365.81) N/mm, was better than that of Gamma nail, which was (224.11±66.45) N/mm; When torsion angle was 5°, torsion strength of control was (1.141±0.191) Nm/du, was better than that of Gamma nail, which was (0.114±0.065) Nm/du. Both press and torsion property were significantly different between two kinds of nail. Conclusion Control is better than that of A-P Gamma nail in both press and torsion property. It may be the main cause that A-P Gamma nail possesses unstability for the treatment of peritrochanteric fracture of femur.
Machine-generated Keywords:
Publication Date:2001-01-01
Online Publishing Date:2025-08-15(First online date of this platform, not the publication date of the document)
Pages:3( 365-367 )
GUANGONG MEDICAL JOURNAL

GUANGONG MEDICAL JOURNAL

PKUISTIC
ISSN:1001-9448
Year, Vol.(Issue):2001,22(5)